Technical Data - Ametek burst NX8 Quick Start Manual And Safety Instructions

Surge, burst, damped oscillatory, telecom surge generators
Table of Contents

Advertisement

AMETEK CTS
Any pressing of a function key will indicate the functions START, CHANGE or CONTINUE. F3 will continue the same test
routine. Also the test time will continue running. If the user selects at first START or CHANGE the test will be stopped
completely.
Page 3 (Change)
CSURGE (1200A) QUICK START
Voltage
Vch
Phase Angle
A
Polarity
+/-
Repetition Rep
:
Trigger
tri
Events
n
:
V
A
+/-
Rep
100
0
+
F1
F2
F3
Press CHANGE and the test parameters can be changed. Select the desired parameter with the related parameter and the
available range is displayed. Change the value by turning the front panel knob. The cursor allows the user to define the digit
to be changed (fast or slow change).
Note: At polarity setting ALT it is necessary to double the number of impulses.
Example n=2 => one impulse positive and one impulse negative.
9.8.

Technical data

Test level
Charging voltage
Short circuit current
Range I
Range II
Range III
Wave shape
Rise time tr
Pulse duration
Polarity
Repetition rate
Events preselection
Counter
Trigger
Trigger of pulses
Synchronization
Resolution
Output
Direct
Test box
HV-coaxial connector; Zi = 2
Measurements
CRO
CRO Û
CRO Î
Peak current meter
:
100V
-
2500V
:
0dgr
-
360dgr
:
+, -, ALT
1s -
100s
:
AUTO, MAN
1 -
30000
tri
n
1
Auto
0001
F4
F5
F6
100 – 2500 V ± 10 %
1'200 A ± 10 %
1 – 18 A
6 – 140 A
40 – 1200 A
8 s ± 20 %
20 s ± 20 %
Pos., Neg., Alt
1 s* - 100 s
1 - 30'000 or endless
1 - 1000000
AUTO, MAN, EXTERN
0 - 360° (16 - 500 Hz)
At the front panel L - PE
Mounted on top of the generator
5 V Trigger
100 mV / V Only for measuring residual voltages up to 500 V on VDR's
Range I
Range II 100 mV/A
Range III 10 mV/A
1200 A
Quick start guide - Single pulse generators
F7
1 V/A
1.00
63 / 88