Safety Circuit For Devices Of 500-Series - Ametek burst NX8 Quick Start Manual And Safety Instructions

Surge, burst, damped oscillatory, telecom surge generators
Table of Contents

Advertisement

AMETEK CTS
Important for operation
Connect the delivered Safety Circuit Terminal SCT
Circuit Adapter SCC AD
The generator does not start any test if the safety circuit is not connected and closed.
NOTE: The short circuit is already short-circuited at the delivered Safety circuit terminal
SCT and Safety Circuit Adapter SCC AD
.
SCT Safety Circuit termination
The safety circuit terminal SCT is part of the delivery and includes the short circuit for the
safety circuit. If an external safety circuit is required, please order the optional Safety
Circuit Adapter SCC AD.
Figure 13 - SCC AD option for external safety circuit loop
3.1.3.

Safety circuit for devices of 500-series

A test can only be started if the safety circuit is closed. By opening the safety circuit
during a running test, the test will be stopped immediately and the high voltage will be
switched Off and grounded.
The safety circuit will not switch off the EUT Mains supply. For switch off the EUT mains
a similar safety circuit must be used as described in a rack solution in the safety
manual..
to the SYSLINK plug.
1.00
Quick start guide - Single pulse generators
or the optional Safety
Figure 12 - Safety and
Warning Lamp
Connectors
Figure 14 - Safety
Termination
13 / 88