Test level
A
Selectable phase angle to superimpose the pulse to the mains (only with external CDN)
+/-
Polarity of the generated pulse
Cpl
Not used in VSS generators
rep (t1)
Repetition rate of the generated pulses
tri
Trigger mode; AUTO, MAN or EXT
n
Number of pulses to be generated
All function keys with exception of F2 (manual Trigger) will stop the test.
Page 4 (Start)
VOLTAGE SURGE
V
= 4000 V
+/-
=
+
rep
=
10 s
n
= endless
Uset = 4000V
U =
STOP
I =
F1
F2
F3
After the Start button was pressed the test voltage indication is blinking. The user can change the blinking value by turning the
digital knob. Any parameter to be changed can be selected by the operator by using the cursor buttons Up and Down.
All functions keys except F2 (MAN TRIGGER) can stop the test routine. The latest setting will be displayed.
Pressing the ESC button will bring the user back to the previous level.
Any pressing of a function key will indicate the functions START, CHANGE, CONTINUE or PRINT. F3 will continue the same
test routine. Also the test time will continue running. If the user selects at first START, CHANGE or PRINT the test will be
stopped completely.
Page 4 (Change)
Surge
Voltage
V
:
Angle
A
:
Events
n
:
Repetition
rep
:
V
A
+/-
coupl
500
0
+
F1
F2
F3
The possible range is displayed for each parameter selected.
The value for change can be selected with the function key and than changed with the INC/DECR knob. By moving the cursor
the digit to be changed can be selected to speed up the procedure.
With ESC the operator can go back to the previous menu and start testing with the new parameters.
8.7.

Technical Data VSS 500N12.x

Capacitor
Energy at 12kV
Z out
Parallel resistor
Open circuit voltage
Rise time tr
Pulse duration
Inside resistor
Polarity
Repetition rate
Events
Counter
Quickstart
A
=
0
grd
cpl
=
/
tri
=
Auto
+ 2000V
Counter
+ 0000A
0000043
F4
F5
F6
Quickstart
250V
-
12000V
0 grd
-
360 grd
1
-
30000
1s
-
100s
rep
n
tri
/
4
00001 Auto
F4
F5
F6
VSS500N12.1
250 nF
18 J
40 
300 
500 V – 12000 V
1,2 s  30 %
50 s /  20 %
Depending on the Ri value, fixed options are 40 , 50  or 500 ,
Pos, Neg, Alternating
1 s* - 100 s
Preselection 1 - 30'000 or endless
1 - 1'000'000
Quick start guide - Single pulse generators
F7
F7
VSS500N12.2
125 nF
9 J
40 
600 
 10 % (general)
 3 % as per IEC 61180 on demand
1.00
VSS500N12.6
250 nF
18 J
50 Ω
300 Ω
55 / 88