Responsibility Of The Operator; General Hazard - Ametek burst NX8 Quick Start Manual And Safety Instructions

Surge, burst, damped oscillatory, telecom surge generators
Table of Contents

Advertisement

AMETEK CTS
Before using this equipment, read the user manual and the
separate delivered safety manual carefully
CAUTION
1.3.

Responsibility of the operator

These operating instructions form an essential part of the equipment and must be available to the operator at all times. The
user must obey all safety instructions and warnings.
The purpose of this instrument is the generation of defined interferences signals for EMI immunity
testing. Depending on the arrangement of the test rig, the configuration, the cabling and the properties
of the EUT itself, a significant amount of electromagnetic radiation may result that could also affect other
equipment and systems.
The equipment is designed to operate in industrial environment. For operating in other or sensitive
environment, such as light industry, airport area..., the user may use a shielded room for operate.
The user himself or herself is ultimately responsible for the correct and controlled operation of the rig.
CAUTION
In case of doubt, the tests should be carried out in a Faraday cage.
1.4.

General hazard

Before applying power to the system, verify that your product is configured properly for your particular application.
The generators and their accessories operate at high voltages.
Hazardous voltages may be present when covers are removed. Qualified personnel must use extreme
caution when servicing this equipment.
Circuit boards, test points, and output voltages also may be floating above (below) chassis ground.
WARNING
The design of external insulation must be such that it exceeds the maximum impulse voltages of the
generator (up to 20 kV in dependency of the generator).
Only qualified personnel who deal with attendant hazards in impulse generators, are allowed to perform installation and
servicing.
Ensure that the AC power line ground is connected properly to the Power Rack input connector or chassis. Similarly, other
power ground lines including those to application and maintenance equipment must be grounded properly for both personnel
and equipment safety.
Always ensure that facility AC input power is de-energized prior to connecting or disconnecting any cable.
The user must ensure that the output power lines are labeled properly as to the safety hazards and that any inadvertent
contact with hazardous voltages is eliminated.
Guard against risks of electrical shock during open cover checks by not touching any portion of the electrical circuits. Even
when power is off, capacitors may retain an electrical charge. Use safety glasses during open cover checks to avoid personal
injury by any sudden component failure.
Neither AMETEK CTS GmbH, nor any of the subsidiary sales organizations can accept any responsibility for personnel,
material or inconsequential injury, loss or damage that results from improper use of the equipment and accessories.
Quick start guide - Single pulse generators
1.00
7 / 88