Fuse for high voltage power supply
10 Safety circuit
11 Warning lamp
9
Fuse F3 of the high voltage power supply
The high voltage power supply unit is protected against overload by this fuse. In case that no pulses are generated anymore
please check the fuse.
10
Safety circuit
The test can only be started if the security circuit is closed. If the circuit is opened during a running test the simulator will be
switched off immediately.
11
Warning lamp
A control signal of (+24V/100mA) is available for external warning indications (warning lamp). The signal is generated after
pressing TEST ON.
12
Reference earth connection
The generator has to be connected to the reference earth plane of the test set up. The connection at the rear part of the
generator is an alternative to the grounding point at the front panel
13
SYNC input
An ac voltage to which the events shall be synchronized is connected to this input. If no voltage is available the tests are
started automatically in asynchronous mode. The maximum input voltage is 250VAC/400Hz.
Quick start guide - Single pulse generators
12 Reference earth connection
13 Sync input