Err 7: Dsp Signal Error; Err 8: Temp Input Error; Err 9: Press In Error; Err 10: Totalizer Overflow - Panametrics Sentinel User Manual

Table of Contents

Advertisement

Chapter 4. Error Codes
4.2.8

Err 7: DSP Signal Error

4.2.9

Err 8: Temp Input Error

Problem:
This message indicates a temperature input error.
Cause:
The temperature exceeds the specified limits for the analog/RTD inputs option card or no input device is
connected.
Action:
Check the temperature transmitter and the connecting cable. Refer to Diagnostics in Chapter 2, Initial
Setup, and recalibrate the analog/RTD inputs option card.
4.2.10

Err 9: Press In Error

Problem:
This message indicates a pressure input error.
Cause:
The pressure exceeds the specified limits for the analog input card, or no input device is connected.
Action:
Check the pressure transmitter and the connecting cable. Refer to "Entering Temperature and Pressure
Constants" on page lxxvii and calibrate the analog input card.
4.2.11

Err 10: Totalizer Overflow

Problem:
The totalizers are unable to keep up with the total accumulated flow signals.
Cause:
The programmed units/pulse value is too small.
Action:
Select a larger number of units/pulse value.
94
Sentinel™ User's Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Sentinel and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents