4.2.4
Err 3: Velocity Range
Problem:
The velocity exceeds the limits programmed in the New Meter Browser information setup (see Chapter 2,
Initial Setup).
Cause:
This error may be caused by the entry of improper programming data or by poor flow conditions and/or
excessive turbulence.
Action:
Make sure the actual flow rate is within the programmed limits. See Chapter 2, Initial Setup for details.
Refer to Chapter 5, Diagnostics, to correct any flowcell and/or transducer problems.
4.2.5
Err 4: Signal Quality
Problem:
The signal quality is outside the limits programmed in the New Meter Browser information setup (see
Chapter 2, Initial Setup).
Cause:
The peak of the upstream or downstream correlation signals has fallen below the correlation peak limit,
as set in the New Meter Browser information setup. This may be caused by a flowcell or electrical
problem.
Action:
Check for sources of electrical interference and verify the integrity of the electronics console by
temporarily substituting a test flowcell that is known to be good. Check the transducers and relocate
them, if necessary. See Chapter 5, Diagnostics, for instructions.
4.2.6
Err 5: Amplitude Error
Problem:
The signal amplitude exceeds the limits programmed in the New Meter Browser information setup (see
Chapter 2, Initial Setup).
Cause:
Solid or liquid particulates may be present in the flowcell. Poor coupling for clamp-on transducers.
Action:
Refer to Chapter 5, Diagnostics, to correct any flowcell problems.
4.2.7
Err 6: Cycle Skip, Accel.
Problem:
The acceleration exceeds the limits programmed in the New Meter Browser information setup (see
Chapter 2, Initial Setup).
Cause:
This condition is usually caused by poor flow conditions or improper transducer alignment.
Action:
Refer to Chapter 5, Diagnostics, to correct any flowcell and/or transducer problems.