Chapter 2. Initial Setup
2.6.1.2
Fluid
Figure 38 on page lvii shows a typical
• Fluid: Choose the fluid from the list box. The system will suggest theoretical soundspeeds. These suggested
soundspeeds can be overwritten by selecting "Other" from the list. Other fluids can be measured by selecting
"Other" and entering the fluid's calculated soundspeed.
• C3 Theory: The theoretical soundspeed of the fluid being measured.
• K Viscosity: The kinematic viscosity of the fluid being measured.
• Const Press: The pressure of the measured fluid. If analog inputs are used for pressure measurement, this box
will not be editable and will show the device number of the analog input.
• Base Press: The base pressure used for standard volumetric measurement calculations.
• Const Temp: The temperature of the measured fluid. If analog inputs are used for temperature measurement,
this box will not be editable and will show the device number of the analog input.
• Base Temp: The base temperature used for standard volumetric measurement calculations.