11.0 ROUTINE MAINTENANCE

THE OPERATIONS OUTLINED IN THIS CHAPTER ARE TO BE
PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL

11.1 Model 300 maintenance schedule

The following are the recommended periodic maintenance items for the Teledyne API
Model 300 CO Analyzer:
Date Instrument Was Recieved: ___________________
J
A
ITEM
N
Particulate
filter element
Pump
diaphragms
Sample flow
Pneumatic
lines
Leak Check
Clean optical
bench

11.2 Replacement of sample filter

a.
Turn off the Analyzer.
possibility of sucking debris into the Analyzer while changing the filter
element.
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 93
NOTE
ONLY!
F
M
A
M
J
J
A
E
A
P
A
U
U
U
B
R
R
Y
N
L
G
TABLE 11.1
This will stop the pump and eliminate the
S
O
N
D
E
C
O
E
RECOMMENDED
P
T
V
C
ACTION
Repalce weekly or as
needed.
Replace every 6
months.
Check for proper flow
(800cc/min ±10%)
annually.
Examine and clean as
necessary.
Leak Check after
maintenance and at
least annually.
As needed.