Figure 7-25 F.B. Tab - FE Diagnostics Dialog Box - RAM Test (RFT Configuration)
3) Check the test results and proceed as follows:
4) Reboot the system.
3
EVISION
If no errors were identified, proceed with the next test.
If no errors were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in the
Calibration
section, on page 6-11, and then repeat the test.
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
and Function
(Theory). If the problem persists, contact your local GE Medical Systems field
engineer for assistance.