6) Under Auto Detect, click the Start button. Checking of the test results commences and the light
adjacent to the button flashes green and continues to flash while Auto Detect is in progress. If a fault
is detected, the flashing light changes to red and an error log is displayed listing the channel/s on
which the test failed.
Alternatively, review the information manually, using the scroll arrows on the right of the Ultrasound
Scope window to select a vector between 0 and 63. The selected vector is displayed in the Vector
# field and its test results are displayed in the Spectrum area.
Check the test results for each vector, as follows:
Note: If the RMS level is above or below the specified level, reboot the system, perform
calibration and repeat the test.
7-24
3
EVISION
Figure 7-13 Ultrasound Scope Window
Verify that the RMS level displayed in the RMS field is above 84 (in absolute value).
Verify that the amplitude at frequency 2.5 is within a range of -5 to -20.
Verify that the amplitude at frequency 5 is below -40.
Verify that the amplitude at frequency 7.5 is below -60.
Section 7-3 - Performing Front End (FE) Diagnostics