Mandatory Site Requirements

The following are mandatory site requirements. Additional (optional) recommendations, as well as a
recommended ultrasound room layout, are provided in section
below).
A dedicated single branch power outlet of adequate amperage (see
meets all local and national codes and is located less than 2.5 m (8.2 ft) from the unit's proposed
location. Refer to the
A door opening of at least 76 cm (2.5 ft) in width.
The proposed location for the unit is at least 0.3 m (1 ft) from the walls, to enable cooling.
Power outlets for other medical equipment and gel warmer.
Power outlets for test equipment within 1 m (3.3 ft) of the ultrasound unit.
Clean and protected space for storage of probes (either in their case or on a rack).
Material to safely clean probes.
In the case of a network option:
Note:
All relevant preliminary network outlets installations at the prepared site must be performed by
authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to
perform servicing of the equipment.
2-3-3

Site Recommendations

The following are (optional) site recommendations. Mandatory site requirements are provided in the
Mandatory Site Requirements
A door opening of 92 cm (3 ft) in width.
An accessible circuit breaker for a dedicated power outlet.
A sink with hot and cold running water.
A receptacle for bio–hazardous waste, for example, used probe sheaths.
An emergency oxygen supply.
A storage area for linens and equipment.
A nearby waiting room, lavatory, and dressing room.
Dual level lighting (bright and dim).
A lockable cabinet for software and manuals.
2-8
3
EVISION
Electrical Requirements
An active network outlet in the vicinity of the ultrasound unit.
A network cable of appropriate length (regular Pin-to-Pin network cable).
An IT administrator who will assist in configuring the unit to work with your local network. A fixed
IP address is required. Refer to the form provided in
that are required.
section, above.