Do not soak or wipe the probe lens with any product listed above (see Warning). Doing so could result
in irreparable damage to the probe. Follow the care instructions supplied with each probe. In order to
disinfect or sterilize probes without damaging them:
Use gas sterilization and legally marked, sterile, pyrogen-free probe sheaths on probes used for
surgery.
Use cold chemical sterilization suitable for plastic medical scanning devices.
Do not immerse a probe beyond the ridge on its case.
Avoid cleaning and storage temperatures over 60°C (140°F).
Warning: Neurological procedures must NOT be performed on a patient with Creutzfield-Jacob disease
because probes that have been used for such examinations cannot be adequately sterilized.
Warning: Neurological and intraoperative procedures must be performed with the use of legally
marketed, sterile, pyrogen-free probe sheaths. Probes used during neurological surgery must NOT be
sterilized with liquid chemical sterilants because of the possibility of neurotoxic residue remaining on the
probe.
10-6-4-2-1
Cleaning and Sterilizing Probes Used Externally
Table 10-12 Cleaning & Sterilizing Procedure for Probes Used Externally
Step
1
2
3
4
5
6
7
10-16
3
EVISION
To protect yourself from blood-borne diseases, wear approved, non-allergic disposable gloves.
Do not immerse the connector or its strain relief in solution. Use ultrasound probe cleaning towels.
Wash the probe and probe cable in warm soap and water solution (below 80°F). Use only mild
non-abrasive soap.
Follow the manufacturer's cold sterilization instructions to disinfect the probe, including concentration
and soaking time; 10 minutes is a typical disinfecting period.
Follow step 3 for a much longer duration to sterilize the probe; 10 hours is a typical cold chemical
sterilization duration.
When complete, thoroughly rinse the washed lens in clear water. Apply sterile distilled water for at
least 60 seconds. Air dry.
Check the lens for cracking and the cable insulation for separation.
Store the probe in its carrying case.
Section 10-3 - Periodic Maintenance Schedule