Electrical Requirements; Vivid™ 4 Power Requirements; Inrush Current; Site Circuit Breaker - GE Vivid 4 BT03 RFI Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
2-2-5

Electrical Requirements

NOTE:
GE Medical Systems requires a dedicated power and Ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
NOTE:
Please note that image artifacts can occur, if at any time within the facility, the Ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-5-1
Vivid™ 4 Power Requirements
Electrical specifications for the Vivid™ 4 monitor and onboard peripherals are as follows:
Table 2-3
Voltage
100V AC
120V AC
220 - 240V AC
2-2-5-2

Inrush Current

Inrush current is not a factor for consideration, due to the inrush current limiting properties of the power
supplies.
Maximum power requirement = 1.2 KVa
100V AC: 8A
120V AC: 8A
220 - 240V AC: 4A
2-2-5-3

Site Circuit Breaker

It is recommended that the branch circuit breaker for the machine be readily accessible.
POWER OUTAGE MAY OCCUR.
CAUTION
The Vivid 3 requires a dedicated single branch circuit. To avoid circuit overload and possible
loss of critical care equipment, make sure you DO NOT have any other equipment operating on
the same circuit.
3
EVISION
Electrical Requirements
Tolerances
±10%
±10%
±10%
Chapter 2 - Pre-Installation
Op. Current
Frequency
8A
50-60 Hz
8A
50-60 Hz
4A
50-60 Hz
V
™ 4 S
M
IVID
ERVICE
ANUAL
2-3

Advertisement

Table of Contents
loading

Table of Contents