Connecting the ECG

The internal ECG is connected into a rectangular-shaped socket on the patient trace (I/O) panel. The patient
trace (I/O) panel is located on the front of the ultrasound unit, as shown in
socket is clearly labelled and color coded, as shown in
1
Footswitch (black)
2
Phono (blue)
3
ECG (green)
4
External ECG (yellow)
The ECG cable is a modular cable consisting of four different cable parts. The main part (trunk) is a
single cable connecting to the unit at one end, and providing a cable splitter device at the other end.
The splitter contains five receptacles, only three of which are used with the Vivid™ 4 ultrasound unit.
Three color coded electrode cables are inserted into the splitter in the appropriate color-coded
receptacles. Each electrode cable hooks up to the appropriate stick-on electrode by a clip-type
connector. The color coding of the electrodes follows one of two standards that are common in different
parts of the world. The cable splitter has a drawing defining the color codes, names and electrode
placements for each of the three cables, as shown in
3-30
3
EVISION
Figure 3-18 Connection Sockets for ECG Cables
Section 3-5 - Completing the Hardware Installation
V
Figure 3-11
Figure 3-18
below.
Figure 3-19
on page 3-31.
™ 4 S
M
IVID
ERVICE
ANUAL
on page 3-16. Each