GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
8-2-4
Rear Cover Replacement Procedures
8-2-4-1
Tools
Use the appropriate Phillips-type screw drivers, as indicated in the rear cover replacement procedures.
8-2-4-2
Preparations
Shut down the Vivid™ 4 ultrasound unit, as described in
8-2-4-3
Rear Cover Removal Procedure
1) Remove the side covers, as described in the
2) Remove the four screws that secure the rear cover in position. Two screws are located on the sides
of the rear cover, as shown below. The other two screws are located in the AC Input Box Unit, as
shown in
3.) Remove the rear cover.
8-2-4-4
Rear Cover Installation Procedure
1) Return the rear cover in its original position.
2) Insert the four screws and secure the rear cover in position. Two screws are located on the sides
of the rear cover, as shown in
Input Box Unit, as shown in
3) Install the side covers, as described in the
3
EVISION
Figure 8-7
below.
Figure 8-6 Rear Cover Screws
Figure 8-7 Rear Cover Screws in AC Input Box Unit
Figure
Chapter 8 - Replacement Procedures
Chapter 3 -
Side Covers Replacement Procedures
Figure
8-6, above. The remaining two screws are located in the AC
8-7, above.
Side Covers Installation Procedure
V
™ 4 S
IVID
ERVICE
Installation.
on page 8 - 4.
on page 8 - 4.
M
ANUAL
8-7