Input AC Voltage Configuration

The internal Vivid™ 4 voltage is 230V AC.
The Back End and Front End Crates operate on 230V AC.
The Peripherals operate on local voltage.
The Monitor operates on 110V AC internal.
6-2-1

Secondary Voltage Configuration

The AC input voltage is factory preset according to specific requirements, as indicated on the label near
the AC input connector. The AC transformer's primary and secondary voltages are set up by using the
appropriate plugs connected in Jumper1 and Jumper2 on the AC Distribution Box, in accordance with
the information in
Table 6-2 Primary and Secondary AC Voltages
Jumper 1
PRIMARY
Jumper 2
SECONDARY
6-2-2

AC Input Cord

The input voltage power cord and plug must meet the local standards and requirements.
6-2
3
EVISION
Table 6-2
below:
Voltage
Range
Plug P/N
Voltage
Plug P/N
Section 6-2 - Input AC Voltage Configuration
100-120V
220-240V
2253054
2253056
100V
120V
2253057
2253058
V
™ 4 S
IVID
ERVICE
220-230V
220-240V
2253059-2
2253060-2