Mitsubishi VCR Installation Procedure

1.) Place the VCR on the left side of the Vivid™ 4 ultrasound unit's top surface (below the control
console - as shown in
room to access the cables at the rear of the VCR.
Video Cassette
Recorder (VCR)
2.) Connect one end of the VCR power cable (P/N: 2269430) to the power outlet on the Vivid™ 4 rear
panel and the other end to the AC power socket on the VCR rear panel - see
page 8-119.
3.) Using the S-video cable (P/N: 2266743), connect one end to the Video OUT socket on the VCR
rear panel (refer to
Audio IN Socket
Audio OUT Socket
Video OUT Socket
4.) Boot up the Vivid™ 4 ultrasound unit by pressing the ON/OFF button on the control console.
3
EVISION
Figure 8-106
Figure 8-106 Mitsubishi VCR Mounted on Vivid™ 4 Scanner
Figure
8-107) and the other to the Video IN socket on the Vivid 3 rear panel.
Figure 8-107 Mitsubishi VCR Rear Panel
Chapter 8 - Replacement Procedures
below) and next to the B/W Printer, if present, allowing sufficient
V
™ 4 S
M
IVID
ERVICE
ANUAL
Figure 8-107
on
AC Power Socket
RS-232C Socket
8-119