GE Vivid 4 BT03 RFI Service Manual page 590

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
10-7-9-4
Data Sheet for Transducer Source Current Leakage
The test passes when all readings measure less than the values shown in
Record all data on the
CAUTION: When power to the unit is ON, never switch the Polarity and the status of Neutral . Be
sure to turn OFF power to the unit before switching them, using the POLARITY switch and/or the
NEUTRAL switch. Failure to comply with this warning may cause damage to the unit!
Table 10-21 Typical Data Sheet for Transducer Source Current Leakage Test
Unit Power
ON
ON
ON
ON
OFF
OFF
OFF
OFF
3
EVISION
Vivid™ 4 Preventative Maintenance Inspection Certificate
Transducer Tested:
Tester
Power
POLARITY
Switch
NORM
NORM
REV
REV
NORM
NORM
REV
REV
Chapter 10 - Periodic Maintenance
Tester
GROUND
or NEUTRAL
Switch
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
V
™ 4 S
IVID
ERVICE
Table 10-15
and
Table
(see page 10-31).
Measurement
M
ANUAL
10-16.
10-29