GE Vivid 4 BT03 RFI Service Manual page 575

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
Section 10-6
Probe Maintenance
WARNING: Any evidence of wear on a probe indicates that it must not be used. Improper
handling may easily damage ultrasound probes. See the Vivid™ 4 User Manual and also refer to
the probe manufacturer's handling instructions, for more details.
Failure to follow these precautions can result in serious injury and equipment damage. Failure
to properly handle or maintain a probe may also void its warranty.
10-6-1
Probe Checks
Table 10-11 Probe Checks
Step
1
2
10-6-2
Probe Handling
All Vivid™ 4 probes are designed and manufactured to provide trouble-free, reliable service. To ensure
this, the correct handling of probes is important and the following points should be noted:
Do not drop a probe or strike it against a hard surface, as this may damage the probe elements and
the acoustic lens, or may crack the housing.
Do not use a cracked or damaged probe. Any evidence of wear indicates the probe must not be
used. Call your field service representative immediately for a replacement.
Perform a visual check of the probe pins and system sockets before plugging in a probe
Avoid pulling, pinching or kinking the probe cable, since a damaged cable may compromise the
electrical safety of the probe.
To avoid the risk of a probe accidentally falling, do not allow the probe cables to become entangled
with, or to be caught in the wheels of the system.
Protect the probe when moving the unit.
Use a soft cloth and warm, soapy water to clean the probe.
Note:
For detailed information on handling endocavity probes, refer to the appropriate supplementary
instructions for each probe.
10-14
3
EVISION
Item
Check there are no traces of gel on the probe holders. If any gel present, clean the probe
Probe Holder
holders with warm water and a damp cloth to remove all traces of gel (soaking may be required
to remove excess gel).
Check all the probes for wear and tear on the lens, cable and connector. Look for bent pins on
the connector and in the connector socket on the unit. Remove any dust from inside the
connector sockets.
Verify that the EMI fingers around the probe connector socket housing are intact.
Probes
Check the probe locking mechanism and probe switch.
Perform a visual check of transducer pins and connector sockets before connecting a probe.
Verify that the system properly recognizes all probes.
Section 10-3 - Periodic Maintenance Schedule
V
™ 4 S
IVID
Description
M
ERVICE
ANUAL

Advertisement

Table of Contents
loading

Table of Contents