GE Vivid 4 BT03 RFI Service Manual page 463

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
4.) Re-connect the mains power cable [B] to the motherboard (beside the ECG location).
5.) Re-connect the mains power cable (4-pin flat connector [C] - Part No. 2252981) to the rear of the
BEP, and fasten with cable ties.
6.) Re-connect the power cable to the Floppy Disk (labelled [D] in
Figure
7.) Slide the PC-VIC & hard disk out sufficiently to be able to re-connect the power cable (labelled [E]
in
Figure
8.) Re-connect the data cable to the PC-VIC (see
9.) Re-connect the power cable to the Hard Disk (labelled [F] in
above.
10.)Return the Ground cable and screw that fastens the PC-VIC in position, as shown in
on page 8-77.
11.)Re-connect the power cables to the media drives (and CD-RW [G]) and MOD [H] .
12.)Carefully place the ECG Module in the front of the BEP, sliding it into position and re-connecting
the power cable [I] to the rear of the ECG Module. (Refer to the
Procedure
13.)Gently slide the power supply up into the final position in the BEP casing.
14.)Check that all cable connections are secured firmly and fasten with additional cable ties, as
necessary.
NOTE:
Make sure that all data cables, memory boards, etc. are securely in position and that nothing has
erroneously become disconnected during the power supply removal procedure.
15.)Stand the BEP2 upright.
16.)At the rear of the BEP2, return and loosely tighten the four screws that fasten the power supply in
position on the BEP, as shown in
17.)Return the BEP front cover as described in the
8 - 56.
18.)Return the one screw previously removed from the top of the side cover, as shown in
on page 8-75.
19.)Fully tighten the four screws at the rear of the BEP (refer to
firmly in position.
20.)Return the BEP2 side covers, as described in the
8 - 54.
21) Carefully return the BEP2 to the original position on the machine chassis, and proceed as follows:
a.) Re-connect all the cables on the rear of the BEP (refer to the step-by-step instructions provided
b.) Return the screws of the lower left and right BEP brackets see
a.) Return the two screws of upper BEP bracket
22.)Replace the Vivid™ 4 rear cover - refer to the
23) Remove the Vivid™ 4 left and right side covers - refer to the
page 8 - 4.
24) Re-connect the mains AC power cable for the mains power wall outlet.
25) Turn ON the main circuit breaker located on rear of the Vivid™ 4.
26) Turn ON the Vivid™ 4 by pressing the ON/OFF button on the Control Console.
8-80
3
EVISION
8-65, above.
8-64) to the PC-VIC; refer also to
on page 8 - 72).
in
Back End Processor Installation Procedure
Section 8-5 - Back End Parts Replacement
Figure
8-65, above.
Figure
8-65, above).
Figure 8-59
on page 8-76.
BEP2 - Front Cover Installation Procedure
BEP2 - Side Cover Installation Procedure
on page 8 - 50).
(Figure
Rear Cover Installation Procedure
V
™ 4 S
IVID
Figure
8-64); refer also to
Figure
8-64); refer also to
BEP2 - ECG Module Installation
step
16) to secure the power supply
Figure 8-41
on page 8-49.
8-41).
on page 8 - 7.
Side Covers Installation Procedure
M
ERVICE
ANUAL
Figure
8-65,
Figure 8-62
on page
Figure 8-58
on page
on

Advertisement

Table of Contents
loading