2392751-100, R
IRECTION
Clean UserDefs
CAUTION: THIS PROCESS IS IRREVERSIBLE! Before proceeding, make sure you have a
complete backup of the current settings if they will be required in the future.
When selecting Clean Userdefs, the following warning message is displayed:
After making a back-up copy of the current settings, click OK to proceed.
NOTE:
This function performs the same procedure as that of Restore Factory Presets (accessed under the
System Configuration window - System tab).
7-116
3
EVISION
-
Figure 7-90 Clear User-defined Settings
Section 7-5 - Common Service Interface
used to clean all user-defined settings and restore the default
settings