Setting Up for Connectivity

Notes:
Four passwords are required to complete the setup of the connectivity options. Ensure that all these
passwords are installed on the Vivid™ 4 ultrasound unit before you begin the configuration
procedure.
The HL7 option (not listed) is a licence granting permission to connect the system to the HL7 gateway,
through the EchoPAC. This function (which is configured on EchoPAC, not on the Vivid™ 4 system)
is used to export examination results to the Hospital Information System (HIS), through HL7 protocol.
Users requiring this option should contact the GE representative.
1) Press Config on the alphanumeric keyboard. The System Configuration window is displayed.
2) Trackball to the System Options tab and press Select. The System Options tab is displayed, as
shown below:
3) Verify that the following options are installed, as required:
3-54
3
EVISION
Figure 3-35 System Options Tab
DICOM Media Creator is required to support DICOM storage to MOD or CD-R media.
DICOM Network Connectivity is required to support DICOM storage over the network, for
example, to Proslov.
EchoNET Connectivity is required to support an EchoPAC direct connection.
Excel Export is required to support the ability to export measurements and reports to Microsoft
Excel files.