GE Vivid 4 BT03 RFI Service Manual page 432

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
Note: When replacing a BEP, one must re-install software using the software provided with the
system. Do NOT use the software supplied with the BEP unless it is a version later
than the software currently installed on the system.
5) Place the BEP on a table and start removing the required parts (listed below), as applicable.
NOTE:
When performing these procedures, handle boards carefully and take the following precautionary
measures:
CAUTION:
TAKING THE NECESSARY ESD PRECAUTIONS:
1.ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD
CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (TO THE RIGHT OF
THE POWER CONNECTOR).
2.FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
For BEP2 (RFT) and (RFI) systems:
PC VIC Assembly - see
ECG Box Assembly (Patient I/O Module) - see
8 - 70.
Plug & Scan Battery - see
PCB: Plug & Scan Card - see
PCB: SCSI Card - see
MO Drive 5.25 - see
3
EVISION
Lower Left
Bracket
Figure 8-41 BEP - Position of Brackets on BEP2 Models
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO
PC-VIC Removal Procedure
Plug & Scan Battery Removal Procedure
SCSI Board Removal Procedure
MO Drive Removal Procedure
Chapter 8 - Replacement Procedures
on page 8 - 73.
BEP2 - ECG Module Removal Procedure
Plug & Scan Board Removal Procedure
on page 8 - 68.
V
™ 4 S
IVID
Upper
Bracket
Lower Right
Bracket
on page 8 - 58.
on page 8 - 57.
on page 8 - 61.
M
ERVICE
ANUAL
on page
8-49