GE Vivid 4 BT03 RFI Service Manual page 484

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
Note:
In order to manipulate and loosen the gas spring cable sufficiently to release and remove
the cable end from the slot in the up/down/swivel handle, it may also be necessary to
loosen the other end of the cable from where it is attached to the cable holder located on
the gas spring rod. If required, refer to the instructions provided in the
Removal Procedure
Similarly, to release the other end of the swivel release cable from the vertical axis welding, refer to
the
11.)Loosen and remove the M6 nut from the handle assembly axis screw - see
Handle mounting bracket
Figure 8-89 Removing the Handle from the Handle Mounting Brackets
12.)Holding the Up/Down/Swivel handle in one hand, extract the axis screw from the handle assembly
(pull in the direction shown in
handle from the handle mounting brackets (fastened to the chassis).
3
EVISION
Grooved Slot on Handle
(shown in detail)
Remove cable end from here
Figure 8-88 Detail of Cable End and Slot on Handle
on page 8 - 89.
Swivel Release Cable Removal Procedure
Axis screw
Chapter 8 - Replacement Procedures
Cable End
Stopper
on page 8 - 94.
Figure
8-89, above) and remove the defective Up/Down/Swivel
V
™ 4 S
IVID
Cable
Gas Spring Cable
Figure 8-89
Handle mounting bracket
M6 Nut
M
ERVICE
ANUAL
below.
8-101