TR4 Boards Replacement Procedure

NOTE:
The procedure below provides the instructions for replacing any of the TR4 boards ( P/N 2253036).
8-4-2-1

Tools

Use the appropriate flat and Phillips-type screw drivers and an ESD wrist band strap as indicated in the
TR4 Boards replacement procedure.
CAUTION:
TAKING THE NECESSARY ESD PRECAUTIONS:
1.ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD
CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (TO THE RIGHT OF
THE POWER CONNECTOR).
2.FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
8-4-2-2

Preparation

Shut down the Vivid™ 4 ultrasound unit, as described in
8-4-2-3

TR4 Boards Removal Procedure

1) Remove the FE Board, as described in the
CAUTION: FRONT END BOARDS ARE HEAVY!
FAILURE TO TAKE PRECAUTIONS WHEN LIFTING , COULD RESULT IN PERSONAL INJURY AND/
OR DAMAGE TO THE SYSTEM.
2.) Place the FE Board on a desk or a stable surface, on top of an ESD mat.
3.) Remove each of the four screws (two from each side) that hold the two brackets that secure the
TR4 boards to the FE Board main plate, as shown in
3
EVISION
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO
Figure 8-32 Location of Screws on Securing Brackets
Chapter 8 - Replacement Procedures
Chapter 3 -
Installation.
Front End Boards Removal Procedure
Figure 8-32
below.
V
™ 4 S
M
IVID
ERVICE
ANUAL
on page 8 - 35.