Validating Communication With The Dicom Server - GE Vivid 4 BT03 RFI Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
3-7-5-1

Validating Communication with the DICOM Server

1) Trackball to the Views tab and press Select. The Views tab is displayed.
2) Trackball to the Dataflow button and press Select. The full list of data flows is displayed, as shown
below:
NOTE:
IMPORTANT INFORMATION ABOUT DATAFLOWS:
The system's default dataflow is Local Archive - Int. HD. In this configuration, the Patient List is in the
local archive and the images are stored on the internal hard drive.
In a dataflow configuration such as Int. HD - MOD 525 (DICOM), images are stored both on the internal
hard drive and on media (MOD). The name DICOM is shown in parenthesis, signifying that on the MOD,
the images are in DICOM format.
No Archive is a dataflow that has no output source, for use in emergencies.
CAUTION: DO NOT change the name of an existing dataflow or attempt to add new services to
a dataflow! In the event that any changes or additions are required, contact the On-line Center.
3) Expand the Local Archive-Int. HD + DICOM Server branch. The service DICOM Server is
displayed.
4) Trackball to DICOM Server and press Select, then trackball to the Check button and press select.
3-62
3
EVISION
Figure 3-45 Views Tab
If the connection is established, a green checkmark is displayed.
If the connection cannot be established, a red cross is displayed.
Section 3-7 - Connectivity Setup