Sony Vcr Removal Procedure - GE Vivid 4 BT03 RFI Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
6.) Re-install the Connectors Panels Cover as described in the
Procedure
7) Reconnect the mains AC power cable from the wall outlet.
8) Turn ON the main circuit breaker located on rear of the Vivid™ 4.
9) Turn ON the Vivid™ 4 by pressing the ON/OFF button on the Control Console.
10.)Under System Configuration, VCR/ECG Tab, select the appropriate VCR mode options. For more
information, see
Note:
Sony VCR - Reset Counter:
If the Reset Counter button is activated while the VCR is in the Stop, Pause, or Record
state, the counter display is reset to zero (0:00:00).
8-8-3-5

Sony VCR Removal Procedure

1.) Remove the Connectors Panels Cover as described in the
Procedure
2.) Release the B/W printer by undoing the fastening belt around it (together with the VCR, if present).
3.) Disconnect the RS-232C cable, the Audio and S-Video IN and OUT cables, and Power cord from
VCR's rear as shown in
4.) Remove the VCR from the left side of the Vivid™ 4 ultrasound unit (taking it out from below the
control console and from beside the B/W printer, if present) - refer to
3
EVISION
on page 8 - 8.
VCR/ECG Tab
on page 8 - 8.
Figure 8-122
Chapter 8 - Replacement Procedures
on page 3 - 45.
on page 8-128.
V
™ 4 S
IVID
ERVICE
Connector Panels Cover Installation
Connector Panels Cover Removal
Figure 8-121
M
ANUAL
on page 8-127.
8-129