Front Cover And Air Filter Replacement Procedures - GE Vivid 3N Pro Series Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
8-2-3

Front Cover and Air Filter Replacement Procedures

8-2-3-1
Tools
Use the appropriate Phillips-type screw drivers, as indicated in the front cover and air filter replacement
procedures.
8-2-3-2
Preparations
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
8-2-3-3
Front Cover Removal Procedure
1) Pull out and remove the air filter, as shown below:
2) Remove the side covers, as described
3) The front cover is secured to the base of the main metal frame by a metal bracket on each side.
Remove the upper screw from each bracket, and loosen the lower screw securing the bracket to
the metal frame, as shown below:
04
EVISION
Figure 8-3 Air Filter
Figure 8-4 Front Cover Bracket Screws
Chapter 8 Replacement Procedures
V
IVID
8-2-2 "Side Covers Replacement Procedures" on page
3N P
/E
S
RO
XPERT
ERVICE
Chapter 3 -
Installation.
Remove
Loosen
M
ANUAL
8-4.