Plug and Scan Test
Table 4-17
1
2
3
4
5
4-10
04
EVISION
Plug and Scan Test
Task
Press the On/Off button on the control console for no
more than 3 seconds.
Turn OFF the system's circuit breaker.
Turn ON the system's circuit breaker.
Press the On/Off Button.
Turn OFF the system's circuit breaker. Turn ON the
system's circuit breaker after entering to Standby mode.
Turn OFF the system's circuit breaker.
Press the On/Off button.
Section 4-3 - Functional Check
V
3N P
/E
IVID
RO
XPERT
Expected Result(s)
The system enters Standby mode.
The system remains in Standby mode, indicated by the
blinking On/Off button LED.
The system returns quickly to the Power-ON state -less
than 30 seconds.
The system enters Standby mode and stays in this
mode. The On/Off button LED is blinking.
The system enters the Power-OFF state. The On/Off
button LED is OFF.