GE Vivid 3N Pro Series Service Manual page 302

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-9-6-2
AC Power Input Test (cont'd)
4) Trackball to the AC Power Input button and press Select. The AC Power Input dialog box is
displayed:
5) Trackball to the Show button and press Select to display the voltage.
NOTE:
For repeat readings, select Loop every and set the reading time cycle.
7-64
04
EVISION
Figure 7-38 AC Power Input Dialog Box
Section 7-3 - Performing Front End (FE) Diagnostics
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
M
ANUAL

Advertisement

Table of Contents