GE Vivid 3N Pro Series Service Manual page 245

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-1-1
Auto Sequence (cont'd)
Auto Sequence commences. While the sequence of automatic testing proceeds, the name of the
test currently in progress is displayed in the Test Name field and the current Loop # and Step # are
shown below. In addition, a flashing colored light below the Passed and Failed displays, indicates
the current test status, as follows:
5.) At any stage, trackball to the Pause button (under Execute Sequence) and press Select, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press
Select.
7.) At any stage during the testing sequence, trackball to the View Report button and press Select to
review the current Auto Sequence Test Report.
8.) When Auto Sequence is complete, trackball to the View Report button and press Select to review
the final Auto Sequence Test Report. This will now list all tests performed and show details of which
tests passed or failed.
9.) Trackball to the Exit button and press Select to close the Automatic FE Testing dialog box.
NOTE:
During Auto Sequence, the system will automatically re-boot at least twice. This is normal and should
not be cause for concern.
04
EVISION
Green
= test passed
Yellow
= test in progress
Red
= test failed
Chapter 7 Diagnostics/Troubleshooting
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
M
ANUAL
7-7