Accessing the MUX Diagnostic Test Options (cont'd)
6) Check the test result message and proceed as follows:
7) Trackball to the Exit button and press Select. If required, repeat this test with a different pulser.
7-56
04
EVISION
If one or more channels were identified as bad, trackball to the Save Report button and press
Select, then contact a GE Medical Systems Service Representative.
If no channels were identified as bad, mark the channels suspected as being bad as follows:
Trackball to each of the channels in the Display area and press select.
Trackball to the Mark bad channel checkbox and press Select.
Trackball to Save Report and press Select, and contact a GE Medical Systems service
representative.
If no errors were identified, proceed with the next test.
If no errors were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in
Section 6-8 "Calibration" on page
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
and Function
(Theory). If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
Section 7-3 - Performing Front End (FE) Diagnostics
V
IVID
6-21, and then repeat the test.
3N P
/E
S
RO
XPERT
ERVICE
Chapter 5 - Components
M
ANUAL