GE Vivid 3N Pro Series Service Manual page 293

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-9-1
Accessing the MUX Diagnostic Test Options (cont'd)
Figure 7-30 Report Dialog Box - Normal Channels Test Low Frequency
The Report dialog box contains the following information:
04
EVISION
Test Result Message: Displays the test result.
TR4 Modules: Displays a graphic representation of the front board (four groups of TR4
sockets). Each group consists of four TR4 sockets, each socket containing four vectors
(channels), as follows:
Each vector is represented by a button (highlighted in green) that displays the number of
the socket to which the TR4 is connected.
Each button is highlighted yellow when selected.
Each button is highlighted red if errors were detected for one or more channels.
Checkboxes: Enable you to manipulate bad channels:
Show Only Bad Channels: Select this checkbox to display only bad channels in the
Display area.
Mark bad channel: Select this checkbox to manually mark the currently selected channel
as bad.
Chapter 7 Diagnostics/Troubleshooting
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
M
ANUAL
7-55