Isolated Patient Lead (Source) Leakage – Lead-to-Ground
10-7-6-1
Definition
This test measures the current which would flow to Ground from any of the isolated ECG leads. The
meter simulates a patient who is connected to the monitoring equipment and is grounded by touching
some other grounded surface.
CAUTION: To avoid damaging the unit, never switch the polarity and the neutral status when the unit
is powered ON. Be sure to turn the unit power OFF before switching them, using the Polarity switch
and/or the Neutral switch.
10-7-6-2
Generic Procedure
Measurements should be made with the Ground Open and Closed, with the Vivid 3N unit ON and OFF,
and with the power line polarity Normal and Reversed. Record the highest reading. For each
combination, the operating controls (such as, the lead switch) should be operated to find the worst case
condition.
.
The connection is at service entrance
or on the supply side of a separately
derived system
H (Black)
POWER
OUTLET
N (White)
G (Green)
Building
Ground
Grounding contact switch (use in both
"open" and "closed" positions)
10-24
04
EVISION
Polarity reversing switch
(use both positions)
Figure 10-4 Test Circuit for Measuring Non-Isolated Patient Leads
Section 10-7 - Electrical Safety Tests
Appliance power switch
(use both "OFF" and "ON" positions)
Appliance
H
N
Internal
G
circuitry
Insulating surface
Current meter
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
Patient lead selector switch (if any)
(activated as required)
Patient connected leads
Between each patient
lead and Ground
H = Hot
N = Neutral (grounded)
G = Grounding conductor