GE Vivid 3N Pro Series Service Manual page 490

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
8-8-1-4-1
Setting the Switches on the Paper Tray
1.) Press the OPEN button in the front of the B/W Video Printer.
2.) Ensure that the switches on the Paper Tray are set as shown in
3.) Load the Paper Roll as described in the Printer's Operating Instructions.
Slide Switch
Settings on the
Paper Tray
Figure 8-103 B/W Video Printer, Slide Switches Location and Settings on the Paper Tray
8-8-1-5
B&W Printer Removal Procedure
1.) Remove the Connectors Panels Cover as described in
Procedure" on page
2.) If optional shelf is present, refer to
139.
3.) Release the B/W printer by undoing the fastening belt secured around it (and around the VCR, if
present).
4.) Disconnect the Remote cable, the B/W Video IN cable, and Power cable from rear of the printer
(refer to
5.) Remove the B/W Video Printer from the right side of the Vivid 3N ultrasound unit (taking it out from
below the control console and from beside the B/W printer, if present) - refer to
Video Printer Installed in the Vivid 3N |(Configuration for UP2950 and 2800 Printers)" on page 8-
118.
8-120
04
EVISION
8-8.
Figure 8-102 "B/W Video Printer Connection Cables" on page
Section 8-8 - Peripherals
8-2-5-3 "Connector Panels Cover Removal
8-8-4-5 "Color Video Printer Removal - Procedure 1" on page 8-
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
Figure 8-103
below.
OPEN Tray button
8-119).
Figure 8-100 "B/W
M
ANUAL

Advertisement