GE Vivid 3N Pro Series Service Manual page 256

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-5
RFT Diagnostic Tests (RFT Configuration)
3) In the RFT tab, verify that the Comment area indicates that the system is ready to perform the test.
4) Trackball to one of the following buttons and press Select:
The relevant LED changes color as follows:
Yellow: The test is in progress.
Green: The test has been completed and no problems were identified.
Red: Problems were identified during the performance of the test.
5) Review the test results in the Comment area and proceed as follows:
7-18
04
EVISION
Select the Board Test button to perform Self Test 1 and Self Test 2 sequentially.
Select the Self Test 1 button to perform Self Test 1 independently.
Select the Self Test 2 button to perform Self Test 2 independently.
If no problems were identified, proceed to the FEC test, as described in
Controller (FEC) Diagnostic Tests (RFT Configuration" on page
If no problems were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in
Section 6-8 "Calibration" on page
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
and Function
(Theory). If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
Section 7-3 - Performing Front End (FE) Diagnostics
V
IVID
(cont'd)
6-21, and then repeat the RFT test.
3N P
/E
S
RO
XPERT
ERVICE
7-3-6 "Front End
7-19.
Chapter 5 - Components
M
ANUAL