TR4 Boards Replacement Procedure

NOTE: The procedure below provides the instructions for replacing any of the TR4 boards ( P/N 2253036).
8-4-2-1
Tools
Use the appropriate flat and Phillips-type screw drivers and an ESD wrist band strap as indicated in the
TR4 Boards replacement procedure.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
CAUTION
NECESSARY ESD PRECAUTIONS:
1.ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD
CAUTION
CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (TO THE RIGHT OF THE
POWER CONNECTOR).
2.FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
8-4-2-2
Preparation
1.) Shut down the Vivid™ 3 ultrasound unit, as described in
8-4-2-3
TR4 Boards Removal Procedure
1.) Remove the FE Board, as described in the
FRONT END BOARDS ARE HEAVY!
CAUTION
WARNING
AILURE TO TAKE PRECAUTIONS WHEN LIFTING , COULD RESULT IN PER-
F
SONAL INJURY AND/ OR DAMAGE TO THE SYSTEM.
2.) Place the FE Board on a desk or a stable surface, on top of an ESD mat.
3.) Remove each of the four screws (two from each side) that hold the two brackets that secure the
TR4 boards to the FE Board main plate, as shown in
04
EVISION
Figure 8-45 Location of Screws on Securing Brackets
Chapter 8 Replacement Procedures
V
3N P
IVID
Chapter 3 -
Front End Boards Removal Procedure
Figure 8-43
below.
/E
S
M
RO
XPERT
ERVICE
ANUAL
Installation.
on page 8 - 44.
8-47