Pencil Probe Test (cont'd)
The Report dialog box includes the following:
04
EVISION
Figure 7-24 Report Dialog Box - Pencil Probe Test (Close)
Test Result Message: Displays the test result.
Pencil Probe Relays: Displays a graphic representation of the eight front board pencil probe
relays, as follows:
Each relay is represented by a button (highlighted in green) that displays the relay
number.
Each button is highlighted in yellow when selected.
Each button is highlighted in red if errors were detected for one or more relays.
Checkboxes: Enable you to manipulate bad channels:
Show Only Bad Channels: Select this checkbox to display only the bad channels in the
Display area.
Mark bad Channel: Select this checkbox to manually mark the currently selected vector
as bad.