Software Installation; Installing Application And Operating System Software - GE Vivid 3N Pro Series Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
Section 3-7

Software Installation

3-7-1

Installing Application and Operating System Software

NOTE:
It is recommended to backup the archive and the settings, as described in the Vivid 3N Pro/
Expert User Manual before commencing software installation (also whenever upgrading the
software).
WARNING: Do not attempt to install software that was not designated for your Vivid 3N Pro/
Expert unit. Attempting to install unapproved software will cause irreversible damage to
hardware and software!
Normally the Vivid 3N system is supplied from the manufacturer with the software already installed. In
some cases, it is necessary to re-install the software (for example, after replacing the Back-end
Processor) by following the software installation procedure shown in
Table 3-22
Step
1
2
3
4
5
6
7
8
9
10
.
04
EVISION
Software Installation Procedure
Action
Press the CD drive <Eject> button and remove any
media if present.
Insert the CD provided with the software kit, and press
the <Eject> button again.
Hold down the <ON/OFF> button on the control
console for 5 seconds to shutdown the system.
Select Full Shutdown.
Hold down the <ON/OFF> button.
Press <Enter>.
Press <Eject> and remove the CD, then press
<Eject> again.
Press <CTRL+ALT+DELETE>
Follow the on-screen instructions.
Chapter 3 Installation
V
3N P
/E
IVID
RO
XPERT
Table 3-22
Expected Result
The CD drive tray ejects.
The CD drive tray will load the software CD.
The shutdown options are displayed on-screen.
The system powers down.
The system powers on and the CD performs an
auto-run. A message is displayed on-screen, prompting
you to press <Enter> to perform software installation.
The CD begins the software installation process.
A message is displayed on-screen prompting you to press
<Eject> and remove the CD.
A message is displayed on-screen prompting you to press
<CTRL + ALT + DELETE> to reboot the system.
The machine performs 4 reboots before completing the
installation. The installation process will take 20 to 50
minutes. When the installation process is complete the
machine will be in regular display mode.
S
M
ERVICE
ANUAL
below.
3-35

Advertisement

Table of Contents
loading

Table of Contents