Safety Reminders

DANGER:
OPENING THE AC GROUND LINE (I.E., METER'S GROUND SWITCH IS OPEN),
DO NOT TOUCH THE UNIT!
WARNING: Two people are required to unpack the unit, as it is heavy. Two people are also
required whenever a part weighing 19kg (35 lb.) or more must be lifted.
CAUTION: If the unit is very cold or hot, do NOT turn ON power to the unit until it has had
sufficient time to acclimate to its operating environment.
CAUTION: To prevent electrical shock, connect the unit to a properly grounded power outlet.
Do NOT use a three-prong to two-prong adapter, as this defeats safety grounding.
CAUTION: Do NOT wear the ESD wrist strap when you work on live circuits where more than
30 V peak is present.
CAUTION: Do NOT operate the unit unless all board covers and frame panels are securely in
place, to ensure optimal system performance and cooling.
When covers are removed, EMI may be present.
WARNING: ACOUSTIC OUTPUT HAZARD
Although the ultrasound energy transmitted from the Vivid 3N ultrasound unit is within
AIUM/NEMA standards and FDA limitations, avoid unnecessary exposure. Ultrasound energy
can produce heat and mechanical damage.
NOTE:
The Vivid3N Pro/Expert User Manuals should be fully read and understood before operating
the unit. Keep the manuals near the unit for reference.