GE Vivid 3N Pro Series Service Manual page 380

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
8-2-6-3
Top Cover Lower Section Removal Procedure (cont'd)
7) Cut the cable tie holding the AC cable to the other cables, and unplug the AC cable from the AC
Outlet in the AC Distribution Box, as shown below:
CAUTION: Do not unplug any cables from below the rear panels, except for those indicated in
this procedure.
8) Viewing the unit from behind, there are two small metal plates at the back of the top cover, each
with two screws. Remove only the outer screw from each plate, releasing the small section of the
collar with its two metal plates attached, as shown below:
8-10
04
EVISION
Figure 8-10 AC Distribution Box Cables
Figure 8-11 Top Cover Collar
Section 8-2 - Cover Replacement Procedures
V
3N P
/E
S
IVID
RO
XPERT
ERVICE