Beamformer (Bf) Diagnostic Tests - GE Vivid 3N Pro Series Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-7

Beamformer (BF) Diagnostic Tests

Figure 7-12 BF Tab - FE Diagnostics Dialog Box (RFT Configuration)
4.) 3) Proceed as follows: Trackball to one of the following buttons and press Select:
-
OR
-
5.) Trackball to the Digital Test 0 button and press Select.
6.) When complete, trackball to the Digital Test 1 button and press Select .
7.) In the BF tab, verify that the Comment area indicates that the system is ready to perform the test.
8.) Trackball to the ADC button and press Select. The beamformer digital test and ADC bit tests are
activated sequentially. The test results are displayed, per vector, in the Ultrasound Scope window,
as shown in
04
EVISION
• Trackball to the Board Test button and press Select
• Perform the Digital Test 0 and Digital Test 1 independently, as follows.
Figure 7-13
below.
Chapter 7 Diagnostics/Troubleshooting
V
3N P
IVID
(cont'd)
/E
S
M
RO
XPERT
ERVICE
ANUAL
Data Flow Map
Test Description Area
Test Area
Comment Area
Tabs
7-23