GE Vivid 3N Pro Series Service Manual page 299

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-9-6
Advanced Board Tests
The advanced board tests check two additional parts of the MUX Board: TEE probe control and
Monitoring, as follows:
-
-
7-3-9-6-1 TEE Probe Control
The TEE probe control test checks the position setting of the probe, as well as its normal operation.
1) Access the MUX tab of the FB Diagnostics dialog box, as described in
Diagnostic Test Options" on page
2) Trackball to the Advanced board tests button and press Select. The Advanced System Tests
options are displayed, as shown in
61.
3) Select the radio button next to the appropriate socket.
4) Trackball to the TEE Probe Control button and press Select. The TEE Probe Test dialog box is
displayed, as shown in
settings.
04
EVISION
TEE probe control enables you to check different positions of the probe; by changing the
probe angle, you are able to verify its normal operation.
Monitoring enabled you to check the HVPS and LVPS voltages, and the AC Power Input, as
well as temperatures.
Figure 7-35 MUX Advanced System Tests Options
Figure 7-36 "TEE Probe Test Dialog Box" on page
Chapter 7 Diagnostics/Troubleshooting
7-50.
Figure 7-35 "MUX Advanced System Tests Options" on page 7-
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
7-3-9-1 "Accessing the MUX
7-62, with all the probe