GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
8-2-1-1
Preparations
Shut down the Vivid 3N ultrasound unit, as described in
DANGER: Electrical hazards exist at several points in the system. Familiarize yourself with all
hazardous voltages and high current levels before removing any of the covers.
DANGER: DO NOT wear the ESD wrist band strap when removing parts from the power supply
unit. Before removing any part of the power unit, turn the power OFF and disconnect the power
cord.
CAUTION: Before removing circuit boards, turn the power OFF and wear the ESD wrist band
strap.
8-2-1-2
Order of Cover Removal
Remove the side covers in the following order:
•
Side Covers, as described on
•
Front Cover, as described on
•
Rear Cover, as described on
•
Connector Panels, as described on
•
Top Cover (lower section), as described on
•
Gas Spring Cover, as described on
•
Bottom Keyboard Cover, as described on
•
Speakers Cover, as described on
•
Control Console Bottom Cover, as described on
•
Control Console Top Cover, as described on
04
EVISION
Chapter 8 Replacement Procedures
Chapter 3 -
page
8-4.
page
8-5.
page
8-7.
page
8-8.
page
8-9.
page
8-13.
page
8-14.
page
8-16.
page
page
8-20.
V
3N P
/E
IVID
RO
XPERT
Installation.
8-17.
S
M
ERVICE
ANUAL
8-3