GE Vivid 3N Pro Series Service Manual page 593

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
10-7-9-4
Data Sheet for Transducer Source Current Leakage
The test passes when all readings measure less than the values shown in
Record all data on the
CAUTION: When power to the unit is ON, never switch the Polarity and the status of Neutral . Be
sure to turn OFF power to the unit before switching them, using the POLARITY switch and/or the
NEUTRAL switch. Failure to comply with this warning may cause damage to the unit!
Table 10-21 Typical Data Sheet for Transducer Source Current Leakage Test
Unit Power
ON
ON
ON
ON
OFF
OFF
OFF
OFF
04
EVISION
Vivid 3N Preventative Maintenance Inspection Certificate
Tester
Power
POLARITY
Switch
NORM
NORM
REV
REV
NORM
NORM
REV
REV
Chapter 10 Periodic Maintenance
Transducer Tested:
Tester
GROUND
or NEUTRAL
Switch
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
Table 10-15
and
(see page 10-31).
Measurement
M
ANUAL
Table
10-16.
10-29