GE Vivid 3N Pro Series Service Manual page 285

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-8-7
Port Test
The Port test enables you to check the general logic control and ports logic.
NOTE:
You must reboot the system after performing the Port test.
1) Disconnect all probes.
2) Access the F.B. tab of the FE Diagnostics dialog box, as described in
Board Assembly Options" on page
3) Trackball to the Port Test button and press Select. The port test is performed and the results are
displayed in the Comments area.
4) Check the test results and proceed as follows:
5) Reboot the system.
04
EVISION
If no errors were identified, proceed with the next test.
If no errors were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in
Section 6-8 "Calibration" on page
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
and Function
(Theory). If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
Chapter 7 Diagnostics/Troubleshooting
V
7-27.
6-21, and then repeat the test.
3N P
/E
S
IVID
RO
XPERT
ERVICE
7-3-8-1 "Accessing the Front
Chapter 5 - Components
M
ANUAL
7-47