GE Vivid 3N Pro Series Service Manual page 284

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-8-6
Pencil Probe Test (cont'd)
7) Check the test result message and proceed as follows:
8) Trackball to the Exit button and press Select. If required, repeat this test with a different pulser.
7-46
04
EVISION
If one or more relays were identified as bad, trackball to Save Report and press Select, and
contact a GE Medical Systems service representative.
If no relays were identified as bad, mark the channels suspected as being bad as follows:
Trackball to the channel in the Display area and press Select.
Trackball to the Mark bad channel checkbox and press Select. Repeat as required to
mark additional channels.
Trackball to Save Report and press Select, and contact a GE Medical Systems service
representative.
If no errors were identified, proceed with the next test.
If no errors were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in
Section 6-8 "Calibration" on page
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
and Function
(Theory). If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
Section 7-3 - Performing Front End (FE) Diagnostics
V
IVID
6-21, and then repeat the test.
3N P
/E
S
RO
XPERT
ERVICE
Chapter 5 - Components
M
ANUAL

Advertisement

Table of Contents
loading

Table of Contents