6. Insert the beveled end of the nebulizer capillary tubing into the other end of the
pump tubing.
7. Attach the capillary tubing to the sample inlet of the nebulizer.

Connecting the Drain Tubing Assembly

Table 5-8.
Part No.
09920186
09920267
09985708
09985735
N0695414
N0695415
To connect the drain tubing assembly, use the following steps and refer to the next
figure.
1. Connect the sleeve (Part No. N0695415) and adapter (Part No. N0695414) to
the spray chamber drain. Thread the barbed fitting (Part No. 09920267) into
the coupler and adapter.
2. Attach the Solvent Flex (PVC) tubing (Part No. 09985735) to the barbed
fitting. Attach the other end of the Solvent Flex (PVC) tubing to the tubing
coupler (Part No. 09920186).
3. Use the other end of the tubing coupler (Part No. 09920186) to connect the
drain tubing to the peristaltic pump.
Drain tubing with 1 mm i.d. (0.04 in.) can also be used by following these steps:
Spray Chamber Drain Spare Parts
Drain Spare Part
Barb Fitting Coupler 3/32 Tubing
Barb Fitting Male 10-32 UNF
PTFE Tubing 0.031 x 0.016
Solvent Flex Tubing, 0.11 inches ID
Drain Adapter
Drain Coupling
Peristaltic Pump
289