Torch Viewing Position Alignment; Performance Checks - PerkinElmer AVIO 500 Customer Hardware And Service Manual

Icp optical emission
Table of Contents

Advertisement

Do not use deionized or tap water in the chiller. Only use coolant fluid
Caution
that includes inhibitor (Part No. N0776200).
Ne pas utiliser déminéralisée ou l'eau du robinet dans le refroidisseur.
Attention
Utiliser uniquement du fluide de refroidissement qui comprend un
inhibiteur (N0776200).
Refer to Chapter 4, Installation for more information on filling the chiller.

Torch Viewing Position Alignment

The torch viewing position alignment procedure is used to set the plasma viewing
position of the spectrometer entrance optics for the highest signal intensity. See
Setting the Torch Viewing Position on page 169.
Perform this procedure when:

Performance Checks

Several checks should be done periodically to check instrument performance.
These include the sodium bullet test to check the nebulizer argon flow, the
Background Equivalent Correction (BEC) test, and Coefficient of Variation (CV)
test. These tests are described in Chapter 6, Troubleshooting.
the instrument is first installed or moved to a new location, or
the torch is removed or replaced, or
the plasma induction plates are replaced (important)
Periodic Checks
203